00732467 Rev. D
Page
101
of
110
Bipolar RF Output Test
NOTE:
A Bipolar Test Adaptor is required in order to connect the bipolar output
receptacle to two test leads (active and dispersive) and to the ESU analyzer.
This connection will allow the technician to properly measure Bipolar RF
Output.
1. Follow steps 1-4 of the Monopolar RF Output Test procedure to set-up the gi4000 ESU
and ESU Analyzer.
2. Select the Bipolar Method menu on the gi4000 ESU touchscreen display.
3. Reference
Figure 13-18
below.
4. Connect the Bipolar Test Adaptor to the Bipolar Receptacle on the front panel of the
gi4000 ESU.
5. Connect the
BLACK
dispersive electrode lead of the Bipolar Test Adaptor into the
DISPERSIVE RECEPTACLE
on the ESU Analyzer.
6. Connect the
RED
active electrode lead of the Bipolar Test Adaptor into the
ACTIVE
RECEPTACLE
on the ESU Analyzer.
7. Reference 00732974 Preventative Maintenance Test Data Summary Sheet
– gi4000
ESU for applicable test settings for bipolar mode.
8. Record the measured value from ESU Analyzer for the bipolar mode at each output
power setting on the 00732974 Preventative Maintenance Test Data Summary Sheet
– gi4000 ESU.
Figure 13-18: Bipolar RF Output Test Diagram