5-10
Performance Tests
SR844 RF Lock-In Amplifier
CH2 Display
once
Set the CH2 display to
θθθθ
(degrees).
Sensitivity Down
once
Select
300 mV
sensitivity.
3) Phase response is checked at various frequencies. For each frequency in the table below, perform
steps 3.1 through 3.2.
Test Frequencies
100 kHz
300 kHz
1 MHz
3 MHz
10 MHz
30 MHz
100 MHz
3.1) Set the RF synthesizer to the frequency in the table.
3.2)
Wait for the SR844
UNLOCK
indicator to turn off.
3.3)
Wait for the CH2 reading to stabilize. Record the CH2 reading (degrees).
4) This completes the phase response test. Enter the results of this test into the test record at the end
of this chapter.
Summary of Contents for SR844
Page 10: ...viii SR844 RF Lock In Amplifier...
Page 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Page 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Page 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Page 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Page 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Page 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Page 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Page 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Page 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Page 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Page 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Page 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Page 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Page 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...