4-14
Gain and Time Constant
SR844 RF Lock-In Amplifier
SETL(?)
The
SETL
command resets the Elapsed Time counter to zero, while the
SETL?
query returns the Elapsed Time as a real number, in units of the current Time
Constant, since either the last
SETL
command or since the
Settle...
key was
pressed. The
SETL?
query does not reset the Elapsed Time counter.