Performance Tests
5-3
SR844 RF Lock-In Amplifier
Getting Ready
Introduction
The performance tests described in this section are designed to verify with a high degree
of confidence that the unit is performing correctly.
Serial Number
If you need to contact Stanford Research Systems, please have the serial number of your
unit available. The 5-digit serial number is printed on a label affixed to the rear panel.
The serial number is also displayed on the CH1 and CH2 displays when the unit is
powered on.
Firmware Revision
The firmware revision code is shown on the Reference display when the unit is powered
on.
Preset
Throughout this chapter it will be necessary to preset the lock-in to a known state, the
factory preset state. To do this, press
Shift
then
Recall
.
Warm–Up
The lock-in should be turned on and allowed to warm up for at least an hour before any
tests are performed. The self-test does not require any warm-up period.
Test Record
Make a copy of the SR844 Performance Test Record at the end of this chapter. Fill in the
results of the tests on this record. This record will allow you to determine whether the
tests pass or fail, and also to preserve a record of the tests.
Necessary Equipment
The following equipment is necessary to complete the performance tests. The suggested
equipment or its equivalent be used.
RF Synthesizer
Frequency Range
10 kHz to 200 MHz
Frequency Accuracy
better than 5 ppm
Amplitude Accuracy
1.0 dB
Amplitude
capable of at least +10 dBm output
Recommended
Marconi 2023
Power Splitter
Mini-Circuits ZFRSC-2050 or equivalent
Feed-Through Attenuators
Mini-Circuits CAT-3 or equivalent
Mini-Circuits CAT-20 or equivalent
Summary of Contents for SR844
Page 10: ...viii SR844 RF Lock In Amplifier...
Page 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Page 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Page 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Page 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Page 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Page 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Page 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Page 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Page 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Page 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Page 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Page 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Page 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Page 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...