Gain and Time Constant
4-13
SR844 RF Lock-In Amplifier
Gain and Time Constant Commands
SENS (?) {i}
The
SENS
command sets or queries the Sensitivity. The parameter
i
selects a
sensitivity from the table below.
i
Sensitivity
i
Sensitivity
0
100 nVrms / -127 dBm
8
1 mVrms / -47 dBm
1
300 nVrms / -117 dBm
9
3 mVrms / -37 dBm
2
1
µ
Vrms / -107 dBm
10
10 mVrms / -27 dBm
3
3
µ
Vrms / -97 dBm
11
30 mVrms / -17 dBm
4
10
µ
Vrms / -87 dBm
12
100 mVrms / -7 dBm
5
30
µ
Vrms / -77 dBm
13
300 mVrms / +3 dBm
6
100
µ
Vrms / -67 dBm
14
1 Vrms / +13 dBm
7
300
µ
Vrms / -57 dBm
AGAN
The
AGAN
command performs the Auto Sensitivity function. This command is the
same as pressing
Shift–SensUp
.
AGAN
automatically sets the Sensitivity of the
instrument.
This function may take some time if the time constant is long. This function does
nothing if the time constant is greater than 1 s. Check the Interface Ready bit (bit
1) in the Serial Poll Status to determine when the command is finished.
CRSV(?){i}
The
CRSV
command sets or queries the Close Dynamic Reserve Mode. The
parameter
i
selects High Reserve (
i=0
), Normal (
i=1
), or Low Noise (minimum
close reserve) (
i=2
).
ACRS
The
ACRS
command performs the Auto Close Reserve function. This command is
the same as pressing
Shift–CloseResrv
.
OFLT(?){i}
The
OFLT
command sets or queries the time constant. The parameter
i
selects a
time constant below.
i
time constant
i
time constant
0
100
µ
s
9
3 s
1
300
µ
s
10
10 s
2
1 ms
11
30 s
3
3 ms
12
100 s
4
10 ms
13
300 s
5
30 ms
14
1 ks
6
100 ms
15
3 ks
7
300 ms
16
10 ks
8
1 s
17
30 ks
OFSL(?){i}
The
OFSL
command sets or queries the Time Constant Filter Slope. The
parameter
i
selects 6 dB/oct (
i=1
), 12dB/oct (
i=2
), 18dB/oct (
i=3
), or
24 dB/oct (
i=4
). This command also sets the No Filter mode, with parameter
i=0.
Summary of Contents for SR844
Page 10: ...viii SR844 RF Lock In Amplifier...
Page 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Page 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Page 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Page 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Page 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Page 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Page 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Page 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Page 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Page 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Page 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Page 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Page 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Page 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...