4-12
Signal Input
SR844 RF Lock-In Amplifier
Signal Input Commands
WRSV(?){i}
The
WRSV
command sets or queries the Wide Reserve Mode of the instrument.
The parameter
i
selects High Reserve (
i=0
), Normal (
i=1
), or Low Noise
(minimum wide reserve) (
i=2
).
AWRS
The
AWRS
command performs the Auto Wide Reserve function. This command is
the same as pressing
Shift–WideReserveDown
.
AWRS
automatically sets the
Wide Reserve Mode of the instrument to the minimum reserve without overload.
INPZ(?){i}
The
INPZ
command sets or queries the Signal Input impedance. The parameter
i
selects 50
Ω
(
i=0
) or 1 M
Ω
(
i=1
).
Summary of Contents for SR844
Page 10: ...viii SR844 RF Lock In Amplifier...
Page 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Page 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Page 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Page 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Page 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Page 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Page 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Page 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Page 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Page 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Page 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Page 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Page 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Page 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...