3-38
Scan - Rel
SR844 RF Lock-In Amplifier
XY READY
R[dBm]
θθθθ
These indicators are on if there are XY or R
θ
Rel Values stored for the current scan
frequency. They do not imply that these Rel Values are being applied to the current
measurement.
Rel Mode
This key toggles
REL MODE
.
In
SCAN MODE
with
REL MODE
on, the instrument automatically recalls the stored
Rel Configuration and Rel Values at
each
scan frequency (if previously stored, as
indicated by
XY READY
or
R[dBm]
θθθθ
)
. The
REL
indicator within the CH1 and CH2
displays is on whenever the recalled Rel Configuration and Rel Values are in effect.
If the current scan frequency does not have any stored Rel Values (as indicated by
XY READY
or
R[dBm]
θθθθ
)
, the current configuration and offsets remain in effect. The
REL
indicator is off in this case.
The stored XY Rel Values are applied as X and Y offsets (adjusted for the current
phase). The stored R[dBm] Rel Value is applied as the R[dBm] offset. The stored
θ
Rel
Value is applied as the Reference Phase (there is no phase offset). Offset indicators
within the CH1 and CH2 displays are turned on as appropriate (as well as the
REL
indicator).
Turning
REL MODE
off does not change the current measurement. The current
measurement configuration
remains
in effect. As long as
REL MODE
is off, stored Rel
Values and Configurations are not recalled at any other scan frequencies.
CLEAR ALL
[Shift–Rel
Mode]
This key sequence discards
all
stored Rel Values and Configurations.
CLEAR ONE
[Shift–Start/
Step]
This key sequence discards the stored Rel Values and Configurations at the
current
scan
frequency.
Any key that
changes the
instrument
configuration
All keys in the SIGNAL INPUT and SENSITIVITY sections as well as
Slope
Up/Down, Ratio, Recall, Preset, Ref Z-In, Ref Source
change the measurement
configuration and turn the
REL
indicators off. This does not affect
REL MODE
.
Start/Step
This key is used to step through the scan frequencies. If
REL MODE
is on, stored Rel
Configurations and Values (if previously stored) will be applied.
Summary of Contents for SR844
Page 10: ...viii SR844 RF Lock In Amplifier...
Page 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Page 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Page 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Page 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Page 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Page 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Page 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Page 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Page 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Page 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Page 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Page 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Page 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Page 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...