Performance Tests
5-17
SR844 RF Lock-In Amplifier
7. Input Noise
This test measures the SR844 input noise.
Setup
Connect a 50
Ω
termination to the signal input.
Procedure
1) Press
Shift
then
Recall
to restore the factory preset instrument settings.
2) Press the keys in the following sequence:
Sensitivity Down
until
the sensitivity is
300 nV
.
CH1 Display
three times
Set the CH1 display to
Xnoise
.
Close Reserve
once until
LOW NOISE
is selected.
3) Wait until the CH1 display stabilizes. Record the value of the CH1 display (nV/
√
Hz).
4) This completes the noise test. Enter the results of this test in the test record at the end of this
chapter.
Summary of Contents for SR844
Page 10: ...viii SR844 RF Lock In Amplifier...
Page 12: ...1 2 Getting Started SR844 RF Lock In Amplifier...
Page 32: ...2 2 SR844 Basics SR844 RF Lock In Amplifier...
Page 60: ...3 2 Operation SR844 RF Lock In Amplifier...
Page 102: ...3 44 Shift Functions SR844 RF Lock In Amplifier...
Page 108: ...4 6 Index of Commands SR844 RF Lock In Amplifier...
Page 144: ...4 42 Example Program SR844 RF Lock In Amplifier...
Page 146: ...5 2 Performance Tests SR844 RF Lock In Amplifier...
Page 150: ...5 6 Performance Tests SR844 RF Lock In Amplifier...
Page 156: ...5 12 Performance Tests SR844 RF Lock In Amplifier...
Page 158: ...5 14 Performance Tests SR844 RF Lock In Amplifier...
Page 162: ...5 18 Performance Tests SR844 RF Lock In Amplifier...
Page 166: ...5 22 SR844 Test Record SR844 RF Lock In Amplifier...
Page 168: ...6 2 Circuitry Parts Lists and Schematics SR844 RF Lock In Amplifier...
Page 246: ...Parts Lists SR844 RF Lock In Amplifier 6 80 Schematic Diagrams...