![Spectracom GSG-5 series User Manual With Scpi Manual Download Page 307](http://html1.mh-extra.com/html/spectracom/gsg-5-series/gsg-5-series_user-manual-with-scpi-manual_1343630307.webp)
Command Syntax
SOURce:SCENario:DUPlicate? <satID>
Parameter
satID
– For a list of satellite identifiers, see "SOURce:ONECHN:SATid?" on page 241.
Example Running
SEND:
SOURce:SCENario:DUPlicate? G3
READ:
9
6.3.3.87 SOURce:SCENario:DURATION
Function
Changes the scenario duration before starting the simulation.
Command Syntax
SOURce:SCENario:DURATION [<mode>,][duration]
Parameters
<duration>
is specified in seconds.
<mode>
could be ONCE/FOREVER/LOOPING.
If only
<duration>
is given, then
<mode>
is ONCE by default.
If
<mode>
is FOREVER,
<duration>
cannot be specified.
Notes
This command changes duration of the currently loaded scenario, but does not change the scen
ario file, so that when you try to edit the scenario, you will see unchanged parameters from the
file.
Examples
SEND:
SOUR:SCEN:DURATION ONCE,3600
Set scenario duration to 1 hour, executed once.
6.3 Command Reference
CHAPTER
6
•
User Manual GSG-5/6 Series
Rev.
26
291