ST2516 Series Operation Manual
41
resistance
measurement pattern as 200Ω.
Query syntax: FUNCtion:IMPedance: RES:RANGe?
Return format: <value><NL^END>
Where, <value> can be:
ST2516
:
20.000E-3, 200.00E-3, 2000.0E-3, 20.000E+0, 200.00E+0,
2000.0E+0, 20.000E+3, 200.00E+3,2. 0000E+6
ST2516A
:
200.00E-3, 2000.0E-3, 20.000E+0, 200.00E+0,
2000.0E+0, 20.000E+3, 200.00E+3
ST2516B
:
20.000E-3, 200.00E-3, 2000.0E-3, 20.000E+0, 200.00E+0,
2000.0E+0, 20.000E+3
The :IMPedance:RES:RANGe:AUTO command sets the range auto selection mode in common
resistance measurement pattern.
The FUNCtion:IMPedance:RANGe:AUTO? query returns to the current range mode.
Command syntax:
ON (1)
:IMPedance:RES:RANGe:AUTO
OFF (0)
Where,
Character 1 (integer:49) means ON.
Character 0 (integer: 48) means OFF.
For example: WrtCmd(“FUNC:IMP:RES:RANG:AUTO ON”); Set the range mode in common
resistance measurement pattern as AUTO.
Query syntax: FUNCtion:IMPedance:RES:RANGe:AUTO?
Return format: <NR1><NL^END>
<NR1> = 1 or 0
The:IMPedance:LPR:RANGe command sets the instrument range in resistance tested at low
voltage measurement pattern.
The FUNCtion :IMPedance:LPR:RANGe? query returns to the current range parameter in
resistance tested at low voltage measurement pattern.
Command syntax: FUNCtion:IMPedance:LPR:RANGe <value>
Where, <value>can be the impedance of the tested LCR component or the specific
range values. Its data format is NR1, NR2, NR3, <value> = 0 to 2000.
For example: WrtCmd(“FUNC: IMP:LPR:RANG 15”); Set the instrument range in resistance tested
at low voltage measurement pattern
as 20Ω.
Query syntax: FUNCtion:IMPedance:LPR:RANGe?
Return format: <value><NL^END>
Where
,
<value> can be
:
2000.00E-3, 20.0000E+0, 200.000E+0, 2000.00E+0