ST2516 Series Operation Manual
43
<NR1> = 1 or 0
1: the offset voltage compensation state is “ON”.
0:
the offset voltage compensation state is “OFF”.
7.1.3
APERture Subsystem Commands
The APERture subsystem commands are mainly used to set the measurement speed and the
average times used in measurement.
The APERture? query returns to the current measurement speed and average times used in
measurement.
Command Tree:
The :APERture command sets the instrument measurement speed.
The :APERture? query returns to the current measurement speed.
Command syntax:
Command syntax: APER <FAST, MEDium, SLOW1 or SLOW2>
For example: WrtCmd(“APERture SLOW1”); Set the instrument self-calibration mode as SLOW.
Query syntax: APERture?
Return format: <FAST
,
MEDium
,
SLOW1 or SLOW2> <NL^END>
The :APERture:AVERage command sets the instrument average times used in measurement.
The :APERture:AVERage? query returns to the current average times.
Command syntax:
Command syntax: APER:AVERage <value>
Where, <value> = 1 to 255
For example: WrtCmd(“APERture:AVER 10”); Set nstrument average times used in measurement
as 10.
Query syntax: APERture:AVERture?
Return format: < NR1> <NL^END>
7.1.4
TRIGger Subsystem Commands
The TRIGger subsystem commands set the instrument trigger source and the delay time after
triggering, and triggers instrument measurement.