ST2516 Series Operation Manual 

45 

 

Where,  ON  Set the measurement delay mode as automatic “ON” 
        OFF  Set the measurement delay mode as automatic “OFF” 
For example: WrtCmd(“TRIG:DEL:AUTO ON”); Set the delay mode as automatic “ON”. 
Query syntax: TRIG:DEL:AUTO? 
Return format: <NR1><NL^END> 
            <NR1> = 1 or 0 

1: the measurement delay mode is automatic “ON”. 

            0:

 the measurement delay mode is automatic “OFF”. 

7.1.5

 

FETCH? Subsystem Commands 

The FETCh? Subsystem commands are used to get the last measurement result and the setup of 

the mode. 

Command Tree: 

 

The [:IMP]? query returns ST2516’s last measurement result to its output buffer.   
Query syntax: FETCh[:IMP]? 
 
The data got from the commands can be divided into three types according to different functions 
and display page. 
1. On the measurement display page, compare display interface, bin display page and statistical 
interface and the measurement function is one-parameter mode (functions are R, T, LPR): 
Data format: <primary parameter>, <system status>; 
 
<primary parameter>= the measured value of the current parameter and the format is NR3 
(when the temperature conversion function is enabled and on the measurement display interface, 

the  primary  parameter  is  the  temperature  incrementΔT),  when  existing  outran

ge or 

measurement error, the returned value is “+9.37”. 
 
System status

  -1   

No data in the buffer zone. 

0    Ordinary measurement data. 

              +1 

Measurement status error. 

 
2.

 

On the measurement display page, compare display interface, bin display page and statistical 
interface and the measurement function is two-parameter mode (functions are R-T

LPR-T): 

Data format: <primary parameter>, <secondary parameter>, <system status>; 
 
<primary parameter>= the measurement of the current primary parameter and the format is NR3 
(when the temperature conversion function is enabled and on the measurement display interface, 

the  primary  parameter  is  the  temperature  incrementΔT),  when  existing  outrange  or 

measurement error, the returned value is “+9.37”. 

Summary of Contents for ST2516

Page 1: ...SOURCETRONIC Qualit ts Elektronik f r Service Labor und Produktion DC Resistance meter ST2516 Manual...

Page 2: ...Introduction to Front and Rear Panels 4 2 1 Introduction to Front Panel 4 2 2 Introduction to Rear Panel 5 2 3 Display Zone 6 2 4 Introduction to Buttons on Front Panel 7 2 4 1 MEAS 7 2 4 2 SYSTEM 7 2...

Page 3: ...T2516 19 3 6 1 Temperature Correction TC 19 3 6 2 Temperature conversion t 20 3 6 3 Types of Temperature Sensor 21 3 6 4 Parameter Setting 22 3 6 5 File 22 3 7 BIN SETUP 22 4 System Setup and File Man...

Page 4: ...rement Accuracy 33 5 3 1 Basic Accuracy for Resistance Measurement 33 5 3 2 Accuracy for Resistance Tested at Low Current Mode 34 5 3 3 Accuracy for Temperature Measurement Pt500 34 5 3 4 Accuracy for...

Page 5: ...1 6 TEMPerature Subsystem Commands ST2516 46 7 1 7 COMParator Subsystem Commands 49 7 1 8 BIN Subsystem Commands 52 7 1 9 STATistics Subsystem Commands 56 7 1 10 SYSTem Subsystem Commands 61 8 Handle...

Page 6: ...instrument Introductions to the improvements of the instrument in performance function internal structure outer appearance accessories packing material etc are subject to change without notice If you...

Page 7: ...ty on DUT its elimination is achieved especially when the resistance changes with the temperature greatly better test results will be achieved With statistic analysis function the instrument can make...

Page 8: ...nt above mentioned 3 For high accuracy do not block the left air vent so as to ensure good ventilation 4 The instrument has been specially designed for decreasing noise jamming caused by the AC power...

Page 9: ...warm up time should not be less than 30 minutes 2 Do not turn on or off the instrument frequently This may cause internal data confusion 1 8 Other Features 1 Consumption 30VA 2 Dimensions W H D 235mm...

Page 10: ...CAUTION INPUT ST2516 DC RESISTANCE METER 1 2 3 4 5 6 7 12 11 9 8 10 Figure 2 1 Front Panel 1 USB interface HOST interface of USB 2 LCD and touch screen 480 272 dot matrix 24 bit 4 3 inch TFT and touc...

Page 11: ...keys 11 Trademark and Model Show instrument trademark and model 12 POWER It is the power switch 2 2 Introduction to Rear Panel Figure 2 2 shows the rear panel of ST2516 1 2 3 4 5 6 DEVICE HANDLER RS...

Page 12: ...nt by it 7 USB Interface PC can remotely control ST2516 series through USB DEVICE 2 3 Display Zone ST2516 series adopts 24 bit 4 3 inch LCD touch screen with a resolution of 480 272 The display screen...

Page 13: ...MEAS to enter into display homepage Selectable functions in this page are shown as follows MEAS DISP COMP DISP BIN DISP STATIS DISP MORE 1 2 MEAS SETUP TC t SETUP ST2516 BIN SETUP SYSTEM SETUP MORE 2...

Page 14: ...units or nails to touch the screen Doing so will cause damage to the screen our company shall not be liable to this case 2 6 Starting Up Ensure the power earth ground wire is grounded plug into a 3 w...

Page 15: ...E Other tools TOOLS Soft keys used to enter into operation pages The following test conditions are displayed in the measurement result condition display area on this page These conditions can be set o...

Page 16: ...istance measurement at low voltage mode Measure and display two types of parameters resistance parameters and temperature parameters ST2516 has 9 DC resistance ranges 20m 200m 2 20 200 2k 20k 200k 2M...

Page 17: ...e key again this key will turn to Save Data OFF and the measurement result will not be saved When switching between ON and OFF once a new CSV file will be created in a U disk The measurement data can...

Page 18: ...MEAS DISP or press MEAS to select the COMP DISP soft key to enter into the COMP DISP page shown as figure 3 2 FILE TOOLS MEAS DISP COMP DISP BIN DISP STATICS DISP MORE 1 2 11 36 54 R 499 76 COMP HI C...

Page 19: ...P NG GD OFF NG The beeper will sound when the measurement result is not consistent with the setting value GD The beep will sound when the measurement result conforms to the setting value OFF Turn off...

Page 20: ...GREEN R 479 67 BIN Figure 3 3 Bin display The following parameters can be set on BIN DISP FILE 1 File Manage Touch this key to enter into the file manage page 2 Screen Copy Touch this key to save the...

Page 21: ...rent from the bin setting the corresponding bin will be displayed in green GD Color OFF When the corresponding bin comparator result conforms to the bin setting there is no display GREY When the corre...

Page 22: ...ttons could be enabled The instrument makes a statistic at every trigger 2 When OFF is selected the statistic function will be turned off and other keys and buttons is available 3 4 3 Parameters of St...

Page 23: ...imum measurement result among all measurement results 12 MinIndex Be used to display the test serial corresponding to the minimum measurement result 3 4 4 TOOLS CLEAR Clear all parameters obtained fro...

Page 24: ...trigger mode BUS BUS trigger mode TRIG DELAY Touch this key the following soft keys will be displayed AUTO MANUAL INPUT NOTE The setting of manual measurement delay time is determined by inputting Ran...

Page 25: ...ted under the current environment temperature will be converted to a resistance value under the user set environment temperature For instance a resistor is tested as 100 under 20 If user sets the temp...

Page 26: ...he start of contact R2 is the resistance after the display is stable K is the variance ration of the environment temperature coefficient when the conductor is at 0 For example When R1 is 200m t1 is 20...

Page 27: ...ature coefficient t varies with environment temperature and material conductivity It is supposed that the temperature coefficient of a material at 20 is 20 its temperature coefficient ct at t will be...

Page 28: ...ou can input the set voltage value 0 2V T1 Touch this key a numeric keyboard will pop up and you can input the set temperature value 100 999 9 V2 Touch this key a numeric keyboard will pop up and you...

Page 29: ...n state as ON or OFF When ON is set the corresponding bin in the display zone will be shown as lighted circle When OFF is set the corresponding bin is shown as a horizontal line TOOLS Touch TOOLS the...

Page 30: ...YSTEM SETUP MORE 2 2 Figure 4 1 System Setup On this page the following functions can be set TouchTone Language Password Bus Mode Baud Rate AC Freq Handler VCC LCD Style and Time 4 1 1 TouchTone The f...

Page 31: ...FY Touch this key to modify the password Steps for modify the password Touch MODIFY a numeric keyboard will pop up then you should input the original password and press ENTER to confirm Input the new...

Page 32: ...s selectable EXT INT Press EXT external power supplyis selected Press INT internal power supply is selected 4 1 8 LCD Style Press LCD Style the following six baud rates is selectable DEFAULT CLASSIC P...

Page 33: ...tion of save recall user can save measurement results and configuration information to ST2516 internal Flash or external U disk meanwhile user can recall data from ST2516 internal Flash or external U...

Page 34: ...516001 CSV 2516002 CSV 2516003 CSV 2516999 CSV 2516001 GIF 2516002 GIF 2516003 GIF 2516999 GIF STA CSV IMAGE E U disk Figure 4 2 file structure in a U disk When using a U disk on ST2516 you should pay...

Page 35: ...isplay files stored in the internal FLASH and the external U disk Touch Exit to exit the file manage page Figure 4 3 External file page Figure 4 4 External file page Four files information will be dis...

Page 36: ...to finish inputting Thus the current settings in all pages will be saved to the file NOTE When storing a file if the inputting file number has already existed the save operation will overwrite the or...

Page 37: ...ay the result Manual Press the TRIGGER button on the panel the instrument will test a DUT once and display the result This mode keeps in waiting mode when it is not used External Test a DUT once and d...

Page 38: ...LOW NOTE t is the measurement waiting time t1 is the data processing time approximately is 5ms More detailed information about time parameters please refer to Chapter 7 Handler Interface 5 1 7 Average...

Page 39: ...ws 5 3 1 Basic Accuracy for Resistance Measurement ST2516 Range 20m 200m 2 20 200 2k 20k 200k 2M Current 1A 100mA 100mA 10mA 1mA 100 A 100 A 10 A 1 A OCV 0 7V 3V Resolution 1 10 100 1m 10m 100m 1 10 1...

Page 40: ...ture range 10 0 to 39 9 40 0 to 99 9 Resolution 0 1 0 1 Accuracy in six months 0 30 Rd 0 5 1 0 30 Rd 1 0 Accuracy in one year 0 45 Rd 0 8 0 45 Rd 1 5 1 accuracy 0 3 measured value 0 5 5 3 4 Accuracy f...

Page 41: ...The reason is that the use of three lines is much more inexpensive and much simpler than that of five lines or six lines which is the biggest advantage of using serial interface for communication The...

Page 42: ...USB interface This connection conforms to USBTMC USB488 and USB2 0 protocols 6 2 1 System Configuration Connect USB interfaces on PC and ST2516 through a USB cable 6 2 2 Install the Driver When ST251...

Page 43: ...ST2516 to the USB interface through a USB cable 6 3 2 Installing Driver Methods of installing USBCDC driver are the same with that of installing USBTMC driver When the installation of driver is finish...

Page 44: ...ical BIN 7 1 1 DISPlay Subsystem Commands DISPlay subsystem commands are mainly used to set the instrument display page The DISPlay query returns to the current page Command Tree The PAGE command sets...

Page 45: ...atistics SYST Set the display page to system setup FLISt Set the display page to internal file list The STATe command is used to set whether the measurement display page has refreshed the display resu...

Page 46: ...T2516 For example WrtCmd FUNC IMP RT Set function as R T Query syntax FUNCtion IMPedance Return Format function NL END If the current function is R T it will returns to RT The IMPedance RES RANGe comm...

Page 47: ...er 48 means OFF For example WrtCmd FUNC IMP RES RANG AUTO ON Set the range mode in common resistance measurement pattern as AUTO Query syntax FUNCtion IMPedance RES RANGe AUTO Return format NR1 NL END...

Page 48: ...r 0 The ADJust command is used to execute or clear 0 ADJ Clear 0 ADJ data Command syntax FUNCtion ADJust CLEAr Execute 0 CAJ operation Command syntax FUNCtion ADJust Returned value 0 or 1 NL END 0 it...

Page 49: ...nd syntax APER FAST MEDium SLOW1 or SLOW2 For example WrtCmd APERture SLOW1 Set the instrument self calibration mode as SLOW Query syntax APERture Return format FAST MEDium SLOW1 or SLOW2 NL END The A...

Page 50: ...ANDLER interface BUS Be triggered by the RS232 interface or the GPIB interface For example WrtCmd TRIG SOUR BUS Query syntax TRIGger SOURce Return format INTernal MANual EXTernal or BUS NL END The DEL...

Page 51: ...on is one parameter mode functions are R T LPR Data format primary parameter system status primary parameter the measured value of the current parameter and the format is NR3 when the temperature conv...

Page 52: ...sending mode of each measurement result 7 1 6 TEMPerature Subsystem Commands ST2516 TEMPerature subsystem commands set the instrument trigger source and the delay time after triggering and triggers in...

Page 53: ...ter Return format Reference Temp Temp Coefficient NL END Where the format and unit of Reference Temp Temp Coefficient are the same as above The TEMPerature CONversion DELTa STATe command sets ON and O...

Page 54: ...o the current input mode of the temperature sensor Command syntax TEMPerature SENSor PT or ANALog Where PT Using PT500 as the input signal of the temperature sensor ANALog Using analog voltage signal...

Page 55: ...Command syntax COMParator STATe ON 1 or OFF 0 For example WrtCmd COMP STAT ON Set the comparator function of the instrument as ON Query Syntax COMParator STATe Return format NR1 NL END NR1 1 or 0 1 th...

Page 56: ...ent comparator function The COMParator UPPer query returns to the current upper threshold Command syntax COMParator UPPer Upper threshold Where Upper threshold 0 to 2 2E 6 NR3 the upper threshold of t...

Page 57: ...nction The COMParator PERCent query returns to the current tolerance Command syntax COMParator PERCent Tolerance Where Tolerance 0 to 99 999 NR2 the tolerance of the comparator function unit For examp...

Page 58: ...instrument comparator interface is ON 0 the current counter function of the instrument comparator interface is OFF The COMParator COUNter CLEAr command is used to clear the counter value of the instru...

Page 59: ...unction The BIN MODE query returns to the current limit mode Command syntax BIN MODE ATOLerance or PTOLerance Where ATOLerance set the bin limit modeas absolute tolerance mode PTOLerance set the bin l...

Page 60: ...Cmd BIN COLOr GD RED Set the color of the result mark as RED when the bin comparator result is GD Query Syntax BIN COLOr GD Return format OFF GRAY RED or GREEN NL END The BIN UPPer command sets the up...

Page 61: ...Resistance 0 to 2 2E 6 NR3 the reference resistance of the specific bin unit For example WrtCmd BIN REF 1 20E 3 Set the reference resistance of bin 1 as 20k Query Syntax BIN REFerence Bin NO Return fo...

Page 62: ...the same as above The BIN RESult command querys the last bin comparator result of the instrument Query Syntax BIN RESult Return format NR1 NL END Where NR1 0 to 7 When the judging result of some bin i...

Page 63: ...uery Syntax STATistics STATe Return format NR1 NL END NR1 1 or 0 1 the current statistics function of the instrument is ON 0 the current statistics function of the instrument is OFF The STATistics MOD...

Page 64: ...tics function of the instrument is ON ingore this command Query Syntax STATistics UPPer Return format the format and unit of Upper threshold NL END are the same as above The STATistics LOWer command s...

Page 65: ...NOTE If the statistics function of the instrument is ON ingore this command Query Syntax STATistics PERCent Return format the format and unit of Tolerance NL END are the same as above The STATistics C...

Page 66: ...50E 01 8 The STATistics COUNt command is used to query the statistics compare results Query Syntax STATistics COUNt Return format HI count NR1 IN count NR1 LO count NR1 measurement error count NR1 For...

Page 67: ...ple WrtCmd SYST BEEP STAT ON Set the save state of the instrument as ON Query Syntax SYST BEEP STAT Return format NR1 NL END NR1 1 or 0 1 the current beeper state of the instrument is ON 0 the current...

Page 68: ...must have been saved otherwise ignore this command The SYSTem EXTVCC command sets the VCC of the instrument Handler interface The SYSTem EXTVCC query returns to the current VCC of the instrument Handl...

Page 69: ...WrtCmd TRG NOTE trg will be valid when INIT CONT ON and trig sour bus commands are enabled The IDN query returns ST2516 s identification string Query syntax IDN Return format manufacturer model firmw...

Page 70: ...trument will trigger a measurement 2 PASS2 The low bin 2 comparison result output signal will be valid 3 FAIL The low bin comparison result output signal will be valid 4 5V Internal 5V power output 5...

Page 71: ...ms t5 end of sampling to control output time 2ms t6 measurement delay time See the setup for measurement delay When OVC is OFF Sampling Time 5ms 50 Hz 5ms 60 Hz Fast 20 ms 50 Hz 16 6ms 60 Hz Medium 11...

Page 72: ...e d the License for Manufacturing Measurement Instruments and its number e Marks for test terminal 9 3 Package The instrument generally wrapped in a plastic bag should be packed in a strong packing bo...

Page 73: ...m the date of shipment You should supply us with the warranty card before you enjoy the free maintenance service This warranty does not apply in the event of misuse or abuse of the product or as a res...

Page 74: ...www sourcetronic com SOURCETRONIC GMBH Fahrenheitstrasse 1 28359 Bremen Germany T 49 421 2 77 99 99 F 49 421 2 77 99 98 info sourcetronic com www sourcetronic com skype sourcetronic...