ST2516 Series Operation Manual
33
5.3
Measurement Accuracy
Checking the measurement accuracy should be taken under the following circumstances:
a
.
Warm-up time should be more than 30 minutes.
b
.
Correctly short the test cables, turn 0 ADJ to ON and perform short calibration
by pressing the touch key or 0 ADJ panel.
The correct short of the test cable is as follows:
5.3.1
Basic Accuracy for Resistance Measurement
ST2516
Range
20mΩ
200mΩ
2Ω
20Ω
200Ω
2kΩ
20kΩ
200k
Ω
2MΩ
Current
1A
100mA 100mA 10mA
1mA
100µA 100µA 10µA
1µA
OCV
0.7V
3V
Resolution
1µΩ
10µΩ
100µΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
100Ω
Accuracy
0.1%+3 0.05%+2
0.05%+2
0.2%+2
Temperature
coefficient
300ppm
100ppm
ST2516A
Range
200mΩ
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
Current
100mA 100mA 10mA
1mA
100µA 100µA
10µA
OCV
0.7V
3V
Resolution
10µΩ
100µΩ
1mΩ
10mΩ
100mΩ
1Ω
10Ω
Accuracy
0.05%+2
Temperature
coefficient
300ppm
100ppm
ST2516B
Range
20mΩ
200mΩ
2Ω
20Ω
200Ω
2kΩ
20kΩ
Current
1A
100mA 100mA 10mA
1mA
100µA
100µA
OCV
0.7V
3V