ST2516 Series Operation Manual
47
Query Syntax: :TEMPerature:CORRect:STATe?
Return format: <NR1><NL^END>
<NR1> = 1 or 0
1: the instrument temperature correction function is “ON”.
0:
the instrument temperature correction function is “OFF”.
The :TEMPerature:CORRect:PARameter command sets the “reference temperature” and
“temperature coefficient” of the instrument temperature correction function.
The :TEMPerature:CORRect:PARameter? query returns to the current “reference temperature”
and “temperature coefficient”.
Command syntax:
:TEMPerature:CORRect:PARameter<Reference Temp.>,<Temp. Coefficient>
Where, <Reference Temp.> = -10.0 to 99.9(NR2) means “reference temperature”, unit:
℃
<Temp. Coefficient> = -99999 to 99999(NR1) means “temperature coefficient”, unit
:
ppm/
℃
For example: WrtCmd(“:TEMP:CORR:PAR 25,3390”); Set the instrument “reference temperature”
and “temperature coefficient” as 25
℃ and 3390 ppm/℃ respectively.
Query Syntax: :TEMPerature:CORRect:PARameter?
Return format: <Reference Temp.>,<Temp. Coefficient><NL^END>
Where, the format and unit of <Reference Temp.>,<Temp. Coefficient> are the same as above.
The :TEMPerature:CONversion:DELTa:STATe command sets “ON” and “OFF” of the instrument
temperature conversion function.
The :TEMPerature:CONversion:STATe? query returns to the current state of the temperature
conversion function.
Command syntax:
:TEMPerature: CONversion: DELTa:STATe <ON(1) or OFF(0)>
For example: WrtCmd(“:TEMPerature:Conversion: :DELTa:STATe ON”); Set the instrument
temperature conversion function as “ON”.
NOTE: Turning the temperature conversion function on will stop the temperature correction
function!
Query Syntax: :TEMPerature:CONversion:DELTa:STATe?
Return format: <NR1><NL^END>
<NR1> = 1 or 0
1: the temperature conversion function of the instrument is “ON”.
0:
the temperature conversion function of the instrument is “OFF”.
The :TEMPerature:CONversion:DELTa:PARameter command sets the “initial resistance” , ”initial
temperature” and ”constant k” of the instrument temperature conversion function.
The :TEMPerature:CONversion:DELTa:PARameter? query returns to the current “initial