ST2516 Series Operation Manual
9
3)
Basic Operation
3.1
<MEAS DISP>
Touch the screen or press down [MEAS], the <MEAS DISP> page will be displayed in the screen
shown as figure 3-1.
Figure 3-1 Measure Display
The following measurement parameters can be set on this page.
Measurement function (
FUNC R-T
)
Resistance range (
RANGE AUTO
)
Measurement speed (
FAST SLOW1
)
Measurement temperature (
TC/∆t
)
File manage (
FILE
)
Other tools (
TOOLS
)
Soft keys (used to enter into operation pages)
The following test conditions are displayed in the measurement result/condition display area
on this page. These conditions can be set on the tool function page of <MEAS SETUP> or
<MEAS DISP> page.
0 ADJ (ON/OFF):
ON: turn on the function of SHORT zero clearing (self-calibration function)
OFF: turn off the function
OVC (ON/OFF): compensation of the offset voltage, i.e. compensation of the measurement
error due to thermoelectromotive force on the DUTs. (ST2516/ST2516A)
3.1.1
Measurement Functions
Measurable parameters on ST2516 are as follows: