ST2516 Series Operation Manual
17
instrument buffer memory.
Hi
: Upper limit value, used to be compared, corresponds to the value of UPPER limit.
Lo
: Lower limit value, used to be compared, corresponds to the value of LOWER limit.
When
p
C
,
p
C K
> 1.33, the working capacity is ideal.
When 1.33
≥
p
C
,
p
C K
> 1.00 , the working capacity is qualified.
When 1.00
≥
p
C
,
p
C K
, the working capacity is insufficient.
---------------------------------------------------------------------------------------------------------------------------------
6)
Hi (num): Be used to add up times that the measurement result exceeds the upper limit
value.
7)
Lo (num): Be used to add up times that the measurement result is less than the lower
limit value.
8)
In (num): Be used to add up times that the measurement result passes.
9)
Max: Be used to display the maximum measurement result among all measurement
results.
10)
MaxIndex: Be used to display the test serial corresponding to the maximum
measurement result.
11)
Min: Be used to display the minimum measurement result among all measurement
results.
12)
MinIndex: Be used to display the test serial corresponding to the minimum
measurement result.
3.4.4
TOOLS
CLEAR: Clear all parameters obtained from last statistic.
TRIGGER: Trigger the instrument to execute a measurement.
3.4.5
FILE
1)
File Manage: Touch this key to enter into the file manage page.
2)
Screen copy: Touch this key to save the current screen page to an external U disk. If no U
disk is found, this function will be disabled.
3)
Save Data OFF: Touch this key, this key will turn to <Save Data ON> and each measurement
result will be continuously saved to the .CSV file in U disk. Before a U disk is inserted, the
save function will be disabled. Press the key again, this key will turn to <Save Data OFF> and
the measurement result will not be saved. When switching between ON and OFF once, a
new .CSV file will be created in a U disk. The measurement data can be analyzed by opening
the file through the computer.
NOTE:
Before switching from statistics interface to other interface, please firstly set the statistics