3-113
DNW-A100/A50/A45
DNW-A100P/A50P/A45P
DIF-42 Board (Side A)
3-10-3. VCO Free-running Frequency
Manually Adjustment
Preparing tools
.
Oscilloscope:
TEKTRONIX 2465B or equivalent
.
Frequency counter: ADVANTEST TR5821AK or
equivalent
.
Analog composite video monitor
n
This monitor is for menu displaying. Be sure to connect
it to VIDEO OUTPUT COMPOSITE 3 (SUPER)
connector.
Preparation
1. Check the setting on the sub control panel.
CHARACTER switch
8
ON
2. Check that the equipment has warmed up.
Before starting the adjustment, warm up the VTR and
other equipment through the power for 10 minutes or
more.
Manual Adjustment
1.
Set and connect the oscilloscope as follows:
CH-2: TP109/DIF-42(F-1), DC 2 V/DIV
GND: E100/DIF-42(F-1)
TIME: 100
µ
s/DIV
2.
Connect the frequency counter’s input to the oscillo-
scope’s CH-2 output.
3.
To enter the maintenance mode, press S1101(G-1) on
the SS-63 board.
4.
Enter A23 : CP VR of the maintenance mode.
5.
Enter the following specified sub menu.
For SDI output:
A231 : SDI ENC VR
For SDTI output: A232 : SDTI ENC VR
For SDI input:
A233 : SDI DEC VR
n
In this time, the adjustment mode is the manual, and
“Manual” is displayed on the video monitor. (In the
time data display area of the VTR, “MANUAL” is
displayed.)
6.
Turn the search dial in FORWARD (
2
) direction to
move the
*
-mark to the line of the adjustment data .
(Displays the item title in the time data display area,
“SDI ENC” for example.)
C
DIF-42
2
A
1-661-135-
MADE IN JAPAN
>EP–GW<
1
3
A
B
E
F
G
4
5
6
7
A No.
E501
E100
TP700
TP108
E101
TP302
TP305
TP304
TP303
E502
TP500
TP501
TP503
TP502
TP504
TP507
TP105
TP106
TP104
TP103
TP107
TP505
E500
TP100
TP506
TP300
TP301
E300
E301
D
TP109
3-10. SDI/SDTI I/O Lines Adjustment (CP-297 Board)
3-10-3. VCO Free-running Frequency Manually Adjustment
7.
Change the data until the indicated frequency on the
frequency counter satisfy the specification .
Specification: 27.00
±
0.10 MHz
n
How to change the adjustment data:
Turn the search dial while pressing the JOG button.
8.
To exit the sub menu, press the MENU button once on
the lower control panel.
n
Go to step 5 in order to perform other VCO free-
running frequency adjustment.
9.
To exit the menu of A23 : CP VR, press the MENU
button once again.
Data save (store the adjusted data)
10. Enter A2F : NV-RAM CONTROL, then execute
“SAVE ALL ADJUST DATA”.
.
Message “Save Complete” is displayed on the video
monitor when this data save is completed normally.
n
When loading the previous data without save the
current adjusted data, execute “ALL DATA PREVI-
OUS”.
11. To exit the maintenance mode, press the MENU
button four times.