66
APPENDIX A - Specifications
SIGMA Manual Number 41018
Revision AR
■
IrDA® Serial Infrared Physical Layer Link Specification v1.4(IrPHY), IrDA Serial Infrared
Link Access Protocol v1.1 (IrLAP) and IrDA Serial Infrared Link Management Protocol v1.1
(IrLMP), IrDA Tiny TP v1.1
■
Wireless – 802.11b, 802.11g
■
EIA-RS-232 levels for Asynchronous Transmit/Receive only (RS232)
Specification
Description
Overall Size (Pump)
With Standard Battery
■
Without IV pole clamp – 5.8” H x 4.2” W x 2.5” D
■
With IV pole clamp – 5.8” H x 6.4” W x 4.7” D
With Wireless Battery Module
■
Without IV pole clamp – 6.3” H x 4.2” W x 2.5” D
■
With IV pole clamp – 6.3” H x 6.4” W x 4.7” D
Weight
With Standard Battery
■
Without IV pole clamp – 25.5 oz ± 1.0 oz
■
With IV pole clamp – 33.5 oz ± 1.0 oz
With Wireless Battery Module
■
Without IV pole clamp – 26.5 oz ± 1.0 oz
■
With IV pole clamp – 34.5 oz ± 1.0 oz
Volumetric Accuracy
Accuracy is based on volume collected over one hour using
compatible Baxter and Hospira Standard IV Sets.
Specified accuracy is maintained on Baxter Standard IV Sets
for up to 96 hours (maximum 12 liters) and on Hospira
Standard IV Sets for up to 72 hours (maximum 9 liters)
Infusion Modes
Continuous primary and secondary, Multi-Step, Cyclic TPN,
Amount/Time
BAXTER
HOSPIRA
0.5 – 1.9 mL/hr
±0.1mL/hr
±0.1mL/hr
2.0 – 800 mL/hr
±5%
±5%
801 – 999 mL/hr
±5%
±10%