![background image](http://html1.mh-extra.com/html/sigma/sigmaspectrum-35700abb/sigmaspectrum-35700abb_operators-manual_1268000035.webp)
29
SIGMA Manual Number 41018
Revision AR
13. Review or reprogram the primary infusion as needed:
Running
■
Press the
REVIEW
softkey to view infusion information and clear volume given mL.
Stopped
■
Press
REVIEW/PROGRAM
softkey to display setup screen and change any value.
■
Press
PROGRAM SECNDRY
softkey to program secondary infusion.
Software version 5.02.03 and earlier
1.
Press the
ON/OFF
button to turn the pump on.
2.
Press
BASIC
.
3.
If the prior setup needs to be erased, press
CLEAR ALL
.
4.
Select primary or secondary bag, and press
OK
.
5.
Select
mL/hr
or use the arrow softkeys to scroll through dose rates, and press
OK
.
6.
Enter the flow rate value, and press
OK
.
7.
Enter the VTBI (Volume To Be Infused) in mL and press
OK
.
8.
Confirm the computed infusion time.
9.
Confirm the Volume Given mL value or press
CLEAR
to erase it.
NOTE:
VTBI counts down to zero, while volume given counts from zero up.
10. Press
RUN
to begin the infusion.
11. Confirm all infusion parameters are as intended.
12. Review or reprogram the primary infusion as needed:
Running and Stopped - Press
REVIEW
softkey to view and or change highlighted value and
clear volume given mL.
Setting up a Secondary Infusion
Software version 6.01.00 and later
1.
Make sure you have prepared the pump for a secondary infusion, using the steps in “Pre-
paring the Pump for a Secondary Infusion” section, beginning on page 20.
2.
Follow the steps for programming the primary bag infusion as described in either “Pro-
gramming the Pump Using the Drug Error Prevention Mode” section, beginning on
page 24 or “Programming the Pump Using the Basic Mode” section, beginning on page 28.