Setting Up the Pump
SIGMA Manual Number 41018
Revision AR
2.
Press the
OPTIONS
softkey.
3.
Select
User Options
from the options list and press
OK
.
Software version 6.03.01 and later
From the Care Area selection screen:
1.
Press the
OPTIONS
softkey.
2.
Select User Options from the menu and press ok.
From the RUN or STOPPED screen:
1.
Press the
OPTIONS
softkey.
The User Options menu has three categories:
■
Alarm Settings
■
Display Settings
■
View Information
Alarm Settings
■
Audio Volume
– The audio volume of the pump has three levels; LOW, MEDIUM, and
HIGH.
■
Audio Tone
– set the type of alarm tone that sounds: Short or Long.
■
Standby Delay (hr:min)
– Set the length of time to keep the pump in standby (or Hold) after
setup of the infusion has completed. Settings are from 00:01 to 99:59 (hr:min). Setting the
value to 00:00, or Infinite, results in an infinite Hold period.
■
Near Empty Alert
– When enabled, a callback alert is generated when less than 30 minutes
of infusion remain. If the initially programmed infusion is less than 30 minutes, the alert
shall not be generated regardless of the setting.
■
Down Stream Pressure Limit
– Pressure levels are selectable from three levels: LOW (6 psi ±4
psi), MEDIUM (13 psi ±6 psi), and HIGH (19 psi ±9 psi).
Display Settings
■
Run Screen Options
– Each of the items in this list may be turned ON or OFF. If enabled,
they are included in the alternating screens that are displayed while the pump is running.
■
Audio level indicator – shows L, M, or H (as selected in Audio Volume)
■
Rate mL/hr – shows the mL/hr infusion rate of the current delivery
■
Dose rate – shows the dose rate of the current delivery
■
mL VTBI – shows the remaining Volume To Be Infused, in mL
■
Time (hr:min) – shows the time remaining in the current infusion, in hr:min format
■
Display Adjust
■
Backlight Level – The display backlight level may be adjusted or turned OFF. The back-
light consumes approximately 400 mW when set to maximum brightness; therefore,
battery life is maximized when the backlight is set to OFF.
■
Setup Options
■
Concentration confirmation – Setting this option to ON results in a prompt that
requires the operator to confirm that the concentration value selected from a list of con-
centrations or a user-entered variable concentration is correct.