![background image](http://html1.mh-extra.com/html/sigma/sigmaspectrum-35700abb/sigmaspectrum-35700abb_operators-manual_1268000073.webp)
67
SIGMA Manual Number 41018
Revision AR
Dose Modes
mL/hr, g/hr, mg/hr, mg/kg/hr, mg/min, mg/kg/min, mcg/hr, mcg/
kg/hr, mcg/min, mcg/kg/min, ng/min, ng/kg/min, Units/hr,
Units/kg/hr, Units/min, Units/kg/min, mUnits/min, mEq/hr,
mEq/kg/hr, mUnits/kg/min, mUnits/kg/hr
Flow Rate
0.5 to 999 mL/hr with 0.1 mL/hr increments from 0.5 to 99.9
mL/hr and 1.0 mL/hr increments from 100 to 999 mL/hr
KVO
MDL drug configured KVO rate (default rate of 1.0 mL/hr if
not configured) or programmed rate (whichever is less)
Total Drug Delivered
0.1 to 9999 mL with 0.1 mL increments from 0.1 to 99.9 mL
and 1.0 mL increments from 100 to 9999 mL
Timekeeping
Real Time Clock, battery backed, 10-year life
NOTE:
Clock is set to GMT.
Logging Memory
■
AC Power Adaptor, low
■
24 hr memory of all set up screens except for multi-
stepand cyclic modes that are maintained permanently
■
Separate pump history log and drug event log
■
10,000 ± event capacity. Once the maximum number
of log entries is reached, the data for each new event
replaces the data for the oldest event (the data for oldest
event is lost)
AC Power
AC Power Adaptor, low profile, covers only one outlet,
Medical Grade (EN60601-1-2):
■
Input: 100V-AC-240V-AC, 50-60Hz/200mA
■
Output: 9V-DC/800mA, short circuit protected
■
cord length 3.0 m (~ 9.75 feet)
Use only SIGMA part number 35714 or equivalent
The SIGMA Spectrum Infusion Pump is classified according to
Medical Electrical Equipment standards as:
■
Class II Equipment
■
Type BF Applied Part
■
Continuous Operation
AC Power Adaptor and Protector
Weight 9.7 oz ±0.5 oz.
Specification
Description