APPENDIX A - SPECIFICATIONS
Drug Error Prevention
“A” pump
■
Includes Drug Library
■
Prevents wrong dose rate mode and dose rate
“C” pump
■
Includes Drug Library
■
Prevents wrong dose rate mode and dose rate
■
Includes Wireless Battery Module for wireless communication
Master Drug Library (MDL)
■
PC based, pharmacy edited and controlled, customized in-house list of all IV and epidural
drugs, along with their safe delivery parameters
■
Care area enable:
■
same name/same concentration drugs to have different dose rate limits
■
pump configurations for maximum; rate, VTBI, patient weight and occlusion level
■
Each drug entry includes the care area, drug name, concentration, dose rate mode, bolus
mode, starting dose rate, soft (able to be exceeded) and hard (not able to be exceeded; an
optional setting) dose rate and bolus limits, volume to be infused (VTBI), primary or second-
ary IV container, and pump screen color.
Drug Library Transfer
■
Transfer from a wireless network connection to a pump using a wireless battery module
■
Transfer from the PC directly to a pump via IrDA®
Standard Gravity IV Sets
Standard gravity IV sets from Baxter or Hospira.
Standards
■
IEC60601-1 including collateral standards; Third Party Notified Body Testing (Reference
Electromagnetic Compatibility Tables)