Programming the Pump
SIGMA Manual Number 41018
Revision AR
9.
View the setup data for any programmed step by moving the highlight (using the up arrow
softkey) to the step indicator bar located at the top of the setup screen and then using the
left and right arrow softkeys to move from step to step.
NOTE:
A one-second delay exists from the time a step is selected and when its
setup data is displayed. This delay is to allow rapid scrolling along the step
bar without updating the screen contents repeatedly and unnecessarily.
■
If the pump is stopped, you can change any setup data by navigating to that step and
pressing
OK
to move to the values that must be changed.
■
If the pump is running, you can view any programmed step by pressing the
REVIEW
softkey; however, no values can be changed with the exception of the Volume Given
value, which can be cleared by pressing the
CLEAR
softkey.
10. Press
RUN
to start the program.
The run screen appears as described in the mL/hr or Dose Error Prevention programming
sections with the addition of a program step indicator. This indicator displays the current
program step in the “Step x of y” format, where x is the current step being delivered and y
is the total number of programmed steps.
11. Check the flow and confirm that all clamps are open, there are no kinks in tubing, and
drops are flowing. If everything is flowing properly, push the
YES
softkey.
12. When the program completes, press the
STOP
key.
The program schedule automatically resets itself and may be restarted without entering/
reentering any setup data.
NOTE:
Always verify current program parameters for each step prior to starting a
new infusion.
The program will be retained indefinitely during power off cycles until reset. To reset the
program press the
RESET
softkey from the program stopped screen.
Clearing the entire multi-step program
To clear the entire program, press the
CLEAR PROGRAM
softkey and answer
YES
to the confirma-
tion screen.
Clearing an individual step
To clear the setup data from any individual step:
1.
Make sure the pump is stopped.
2.
Press the
REVIEW
softkey
3.
Use the arrow softkeys to move the highlight to the desired step in the step bar.
4.
Press the
CLR STEP
softkey.
NOTE:
Clearing a step does not delete that step unless it is the last step in the
program.