5
SIGMA Manual Number 41018
Revision AR
CAUTION:
Service Personnel Must be Trained at SIGMA
Servicing Spectrum pumps is restricted to qualified, SIGMA-trained, service
personnel who employ SIGMA authorized parts and procedures. Use of other parts
and servicing procedures is prohibited.
CAUTION:
Perform Preventative Maintenance Annually
Pumps should be tested for proper performance annually and also whenever
damage from drops, fluid intrusion and other causes is suspected. See SIGMA
Spectrum Service Manual for complete information.
CAUTION:
Do Not Improperly Clean Pumps
During cleaning, do not allow fluid to seep inside pump (especially through front
panel door latch holes or back case speaker holes) or severe damage may occur.
Wipe on minimal amounts of cleaning fluids, never spray them. Use only SIGMA
specified compatible cleaning fluids. Do not autoclave or ETO sterilize pumps.
CAUTION:
Be Cautious Near RF Sources
The Spectrum pump meets the electromagnetic compatibility (EMC) requirements
as specified in the International Electrotechnical Commission’s (IEC) 60601-1-2
(2001-09) standard for emissions and immunity. It is good practice to keep the
pump separated away from other equipment, such as hand-held transmitters,
cellular phones and electrosurgical equipment that may generate strong radio
frequency interference (RFI). Refer to the EMC Immunity Section, Separation
Distance, in this manual for recommended minimum distance.
CAUTION:
Confirm Audio Operation
When pressing the ON key and all other keys, confirm that an audio beep is heard.
If sound cannot be heard, discontinue use of the pump and return to SIGMA for
service.
CAUTION:
Confirm Display Operation
Regularly observe the pump’s display. Discontinue use of the pump and return to
SIGMA for service if display abnormalities are observed.
CAUTION:
Electric Shock Hazard
Refer servicing to qualified service personnel at your institution or return to
SIGMA.