![background image](http://html1.mh-extra.com/html/sigma/sigmaspectrum-35700abb/sigmaspectrum-35700abb_operators-manual_1268000024.webp)
18
Operational Overview
SIGMA Manual Number 41018
Revision AR
Secondary Infusions
1.
Lower the primary bag at least 20" below the secondary bag.
2.
Stop the pump if it is running.
3.
Press the
REVIEW/PROGRAM
softkey.
4.
Press the
PROGRAM SECONDARY
softkey.
5.
Select the drug for the secondary infusion.
6.
Push
OK
to select/confirm the secondary bag.
7.
Enter all required values on the setup screen.
8.
Push
RUN/STOP
to begin secondary infusion.
9.
Check the flow and confirm that drops are falling in secondary and not the primary drip
chamber.