37
SIGMA Manual Number 41018
Revision AR
Manual /Multi-Step Programming
The Manual (RAMP/Multi-Step) Programming Mode allows the pump to be programmed with up
to 10 individual infusion steps using either the Drug Error Prevention or Basic Programming Opera-
tions. Drugs are eligible for use in the Manual Programming Mode, provided that the selected drug
has not been specifically assigned to either the primary or secondary bag as identified in the pump’s
Master Drug Library. In software versions 6.04.01 and later “Primary Only, Multi-Step” in the
MDL is also available for multi-step programming.
1.
Press the
ON/OFF
button to turn the pump on.
2.
If the prior setup needs to be erased, press
YES
soft key.
3.
Use the arrow keys to select the care area (nursing area) and press
OK
.
4.
If the drug is to be delievered using the DEP
, select the drug from the MDL as described in
“Programming the Pump Using the Drug Error Prevention Mode” section, beginning on
page 24.
5.
If the drug is to be delivered using the Basic mode,
enter the Basic mode:
■
For version 5.02.03 and earlier, press
BASIC.
■
For version 6.01.00 and later, type the lettters
BA
, select
Basic
, and press
OK
.
6.
On the screen where you select the primary or secondary bag, press the
RAMP/MULT-
STEP
softkey to enter the Manual Programming Mode.
■
Ramp
softkey for software version 5.02.03 and earlier
■
Multi-Step
softkey for software version 6.02.00 and later
NOTE:
The Manual Program Mode is not available for the secondary bag or the
primary bag when a secondary program exists in memory.
NOTE:
If the selected drug has been configured in the MDL as a Primary Only
Multi-step infusion, bag selection will be bypassed and the Manual
Program Mode setup screen will be displayed.
7.
In the Manual Programming Mode, continue with the setup as described in the Basic or
Dose Error Prevention programming sections.
A step indicator bar is located at the top of the screen. The bar shows which steps within
the program have parameter data (a small white highlight) and which step is currently
being viewed (a full white highlight).
8.
When the setup of an individual step has been completed, press
OK
to advance and pro-
gram the next step.
When the 10th step has been programmed, the program schedule is complete and no more
steps may be programmed.
NOTE:
Only one step is necessary to start a program however it must be the first
and only programmed step. The pump may not be started if setup data is
missing from any step in the program. Any parameter data missing within
the program shall be identified in a popup message when a program
attempts to start.