Setup with the integrated HMI
7.8.4 Resets
7.8.4
Resets
Restart transmitter
Is used to restart the device without switching off the supply voltage.
Note
The simulation is interrupted. Saved configurations are not reset
Setting
Selection
Selection
Description
CANCL
Cancel
RESET
DEVICE RSTRT
OK
Restart transmitter
Reset
Is used to provide various reset options of the device.
If the "
Restore factory settings
"
option is selected, all parameters are reset to
the
default settings, except
•
Device address
remains unchanged
• Value of the "
User PIN
"
(write protection) is not reset
•
Peak values
are not reset
•
Automatic false echo suppression
is reset to the default setting (deactivated),
but the detected TVT is not lost
•
User-specific TVT setting
is reset to the default setting (deactivated), but
"
User-specific TVT index markers"
are not lost
Selecting the "
Restore ordered configuration
"
option restores the device settings
ordered by the customer. The parameters that were not configured via the order are
reset to their default settings.
Setting
Selection
Selection
Description
NO
Cancel
FACT
Restore default settings
RESET
RESET
CUST
Restore ordered
configuration
Reset peak values
Resets all recorded peak values.