Setup with the integrated HMI
7.3 Menu overview
Menu item
Parameter
Setting
Default
FAULT CUR =
Main Current
Output_Configuration1:
Current Limits
≤ 3.6 mA, ≥ 21 mA
≤ 3.6 mA
FAIL SAFE LOE = Echo
Loss Detection: Mode
HOLD = off
FAULT = Failure
HOLD
LOE TIMER
Echo Loss Detection: Echo Loss Time
15 s
LINR
Linear
CONIC
Conical Bottom
FLAT
Inclined bottom
CYLIN
Horizontal cylindrical vessel
VESSEL SHAPE
USER
User programmable
LINR
VESSEL DIM A
Intermediate Height
0 mm
VOL UNITS
Level Scaling: Scaling Unit
l
VOLUME
UPPER SCALNG
Level Scaling: Max Scaled Value
100 l
PBFLM
Flow Palmer-Bowlus flume
RWRC
Venturi-, trapezoidal weir,
rectangle weir
TPVWR
Flow VNotch-, triangle weir
PRIM MEASDEV
CUSTM
User programmable
PBFLM
VFLOW UNITS = Level
Scaling: ScalingUnit
l/s, l/min, l/h, MI/d, m
3
/s, m
3
/min, m
3
/h, m
3
/
d, lb/s, lb/min, lb/h, gal/s, gal/min, gal/h,
gal/d, Mgal/d, ft
3
/s, ft
3
/min, ft
3
/h, ft
3
/d, bbl/
s, bbl/min, bbl/h, bbl/d, ImpGal/s, ImpGal/
min, Imp/Gal/h, ImpGal/d
l/s
SCALE FORMAT
Level Scaling: Decimal Places Display
0
VOLUME
FLOW
UPPER SCALNG
Level Scaling: Max Scaled Value
100 l/s
UPPER SCALNG
Level Scaling: Max Scaled Value
100 l
CUSTOM
CUSTOM CURVE =
Linearization Points:
Level Value, Volume
Value
X VALUE 1, Y VALUE 1, X VALUE 2, Y VALUE
2, … X VALUE 32, Y VALUE 32
0 %,
0 … 100 %,
100 %
LOCL DISPLAY START VIEW
Display Config: Picture Selection
LEVEL
Menu item
Parameter
Setting
Default
CONFIDENCE
Level Reliability Value: Unit
ECHO SIG
STR
Level Echo Info: Amplitude
SIG QUALITY
NOISE
AVG
Noise Detection Info: Noise
Level Abs
CONFIDENCE
NEAR
RANGE
Measuring Range
ON
Active
ECHO CONFIG
OFF
Inactive
OFF
SIGNAL
TVT CONFIG
AFES ON False Signal Calculation: Manual
Curve active
ON
32
SITRANS LR150
Operating Instructions, 08/2020,
A5E50352155