Display messages
2
Dis-
played
messages
Mode of
operation
Cause
Possible reasons and advice
Err 3*
During
exposure
Interrupted expo-
sure because of
overvoltage > 142 kV
Only with "Err 3".
•
Check the filament heating value
for the selected kV (see section
"P07 - Read-out filament values
from battery backed RAM").
Compare the value with the value
in the delivery protocol. If the
difference is large, make a new
adaptation (see section "P10 - Fast
adaptation").
"Err 3" in combination with:
•
"Err 21" & "Err 22".
Check the measuring point
"FIL_VAL" on D1 (see page 160**
and page 110**).
•
"CAL 13". Check the 60 V supply
voltage on D4 (see page 160**).
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1