MOBILETT Plus HP
SPR8-220.061.03
Page 2 of 2
Siemens AG
Rev. 03
09.03
CS PS 24
Medical Solutions
3 - 2
Troubleshooting chart
* S27 is the exposure release switch. First step = PREP, second step = EXP.
s27* PREP.
Rotating anode runs
s27* EXP.
Anode speed
reached
High voltage
"Exposure"
Exposure
interrupted
Correct
density
MOBILETT Plus HP
OK
Yes
Yes
Yes
Yes
No
Yes
Exposure release switch,
cables
and connections
Err 6 (see P02 - Rotation test)
Remote control (option)
Exposure release switch,
cables and connections
Displayed message...
Remote control (option)
kV-mAs-selection
Exposure time
Heating value
Darkroom
No
USE 39
Err 6
Err 6
No
No
No
Displayed message...
No
Displayed message...
Yes
Displayed
message
Displayed
message