Replacement and repairs on the collimator
9 - 9
Siemens AG
SPR8-220.061.03
Page 9 of 10
MOBILETT Plus HP
Medical Solutions
Rev. 03
09.03
CS PS 24
•
If the deviation is more than 1.7%, loosen the Allen screws marked "E" slightly and move
the rotary-flange to the desired position.
•
Tighten the screws marked "E" (Fig. 15).
•
Repeat the test and adjustment of the rotary flange position until the deviation between
the light and radiation fields is within the tolerance specified.
•
Reinstall covers "A" and "B".