MOBILETT Plus HP
SPR8-220.061.03
Page 10 of 28
Siemens AG
Rev. 03
09.03
CS PS 24
Medical Solutions
10 - 10
Mechanical systems
•
Tighten screws (2) and (3).
•
Tighten the adjustment screw until the arm and single tank are held in the uppermost
position by the spring force alone.
The installation of the turn plate is performed in the reverse order.
•
Install the turn plate.
•
Reattach the shafts.
•
Reattach the bearing rings.
•
Reattach the locking rings.
•
Connect the ground cable and K68.
•
Reinstall the supporting wheels.
•
Remove the ceiling support carefully and check that the single tank remains in the
uppermost position by itself. If not, tighten the adjustment screw.
•
Move the arm into the park position.
•
Lower the unit to the floor and block the rear wheels.
•
Check the arm and single tank movement. If necessary, adjust according to section
"Adjustment of the arm system".
•
Reattach all covers.
•
Check the function of the turn plate by pressing one of the turn plate buttons and turning
the unit. Check also that the turn plate returns to the park position when the button is
pressed a second time.
•
Check the unit by releasing an exposure.