Service programs
5 - 3
Siemens AG
SPR8-220.061.03
Page 3 of 8
MOBILETT Plus HP
Medical Solutions
Rev. 03
09.03
CS PS 24
If an error is detected, the test program will be aborted, a beep will be signaled and a
message will be displayed:
If the test ends successfully, the display shows:
P03 - Capacitor bank and charging test
5
(Wiring diagram, page 130)
When the test is started the display will show:
During the test, the capacitor voltage is displayed except when the capacitance is shown.
The capacitor will first be charged to 350 V, then discharged to 229 V.
The voltage will be held for 5 seconds before it is restored to 350 V. During the test, the
voltage in the capacitor bank is indicated, except when the capacitance is displayed:
If an error is detected, the test program will be aborted. A beep will be audible and a mes-
sage will be displayed:
If the test ends successfully, the display will show:
Duration: approx. 1.5 minutes.
kV
mAs
P02 Err 6 = Rotation error
kV
mAs
P02 PASS = Rotation test passed in program P02
kV
mAs
CAP tESt = Capacitor test
kV
mAs
CAP 350
= 350 V
kV
mAs
CAP 0.13 F = Capacitance 0.13 Farad
kV
mAs
P03 Err"xx" = See actual "xx" in separate displayed message list
kV
mAs
P03 PASS