HT and single tank error diagnosis
6 - 3
Siemens AG
SPR8-220.061.03
Page 3 of 12
MOBILETT Plus HP
Medical Solutions
Rev. 03
09.03
CS PS 24
Overview
Faulty curve
Possible causes
What to do
Inverter current normal
HT+ discharges
Defective tube
Defective high voltage capacitor (+)
Arching to ground on plus side
Replace single tank
Inverter current normal
High speed HT - discharges
Defective tube
Defective high voltage capacitor (-)
Defective filament transformer
Defective insulation on the wires to
the cathode
Arcing to ground on minus side
Replace single tank
Inverter current normal
High speed HT+ and HT-
discharges
Defective tube
Defective high voltage capacitor (-)
Arcing between plus and minus
side
Replace single tank
Abnormal inverter current;
e.g. triangle shaped HT-dis-
charges
Defective high voltage transformer
Defective diode strip
Defective tube
Replace single tank
Abnormal inverter current
e.g. asymmetric
Defective inverter
Replace inverter
Very low tube current
(normal inverter current) (Err
5)
Connection K64/1 and K64/2
reversed between D13 and single
tank
Properly connect K64/
1 and K64/2
Full output voltage not
obtained after adaptation
Defective high voltage transformer
Replace single tank
Rapid series of voltage
peaks (increasing voltage)
Arcing in HT measurement resistor
or series resistor
Replace single tank