11 - 1
Siemens AG
SPR8-220.061.03
Page 1 of 8
MOBILETT Plus HP
Medical Solutions
Rev. 03
09.03
CS PS 24
Motor drive units
11
Adjustment of the speed
11
•
Adjust the speed with a screwdriver as shown in (Fig. 1).
•
Clockwise rotation decreases the speed.
•
Counterclockwise rotation increases the speed.
Fig. 1
High Voltage!
Life-threatening electric shock hazard exists. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
There is only one backward speed.
MOB00478
CCW =
CW =
faster
slower
MOB00477
DANGER
NOTE