2-44
ACUSON
Freestyle User Manual
Safety and Care
ACUSON Freestyle Rollstand Configuration
WARNING:
When operating and transporting the ultrasound system using the ACUSON Freestyle
Rollstand, insure that the rollstand is configured according to the specifications provided
in this manual. Failure to do so can cause the system to tip over and create a risk for
injury to the user or patient and damage to the system.
5
7
2
4
6
1
3
Rollstand Configuration
1. Quick Release Adaptor Plate: the top of the
plate
should be placed no higher than
48 in/122 cm. Do not place anything other than the Freestyle, probe batteries, and probes
on this mount.
2. Handle: the top of the Handle should be placed no higher than 40 in/101 cm from the
floor.
3.
Keyboard Shelf: should be placed no higher than 33 in/84 cm from the floor.
4. Rear Storage Basket: the top of the basket should be placed no higher than 30 in/76 cm
from the floor. Weight limit 4 lbs/1.8 kg.
5. Front Storage Basket: top of basket should be placed no higher than 24 in/61 cm from the
floor. Weight limit 4 lbs/1.8 kg.
6. Cable Wrap: should be placed no higher than 22 in/56 cm.
7.
Video Printer Shelf: the shelf should be placed no higher than 12 in/30 cm from the floor.
Weight limit 6 lbs/2.7 kg. Secure printer with straps.
Summary of Contents for ACUSON Freestyle
Page 4: ...iv ACUSON Freestyle User Manual ...
Page 14: ...1 2 ACUSON Freestyle User Manual Introduction ...
Page 70: ...3 2 ACUSON Freestyle User Manual Setup ...
Page 78: ...3 10 ACUSON Freestyle User Manual ...
Page 80: ...4 2 ACUSON Freestyle User Manual Patient Studies ...
Page 92: ...5 2 ACUSON Freestyle User Manual Scanning ...
Page 110: ...6 2 ACUSON Freestyle User Manual Wireless Probes ...
Page 122: ...7 2 ACUSON Freestyle User Manual Batteries ...
Page 128: ...8 2 ACUSON Freestyle User Manual Connectivity ...