Wireless Probes
Average Noise Level Indicator
An additional indicator of wireless imaging quality, called the
Average Noise Level
(ANL)
has been calculated for various conditions. ANL is another tool to guide wireless scanning. It
provides a measure of the potential for a compromised wireless link to introduce noise artifact
into the image. ANL is displayed in real-time above the wireless signal quality meter.
ANL is displayed in dB. An ANL value of 0.0 represents the optimal setting and indicates that
the signal quality is very good and free of errors that might introduce noise into the image.
Be alert to the ANL level, in accordance with the guidance below. To insure optimal wireless
imaging, use the information from both the wireless signal quality meter and the ANL.
PATIENT NAME
PATIENT ID
XXX
XXX
XXX
XXX
XXX
XXX
XXX
XXX
L13-5
Nerve Gen
2.5cm
G 9
CG 10
SC 3
SF 6
PP 10
55dB
CF 1
CP 1
Vel 1
Pri 4
20 Hz
ANL 0.0 dB
XXX
XXX
XXX
XXX
--------
ANL 0.0 dB
--------
ANL 0.5 dB
--------
ANL 1.3 dB
--------
ANL 2.9 dB
--------
--------
No Signal
OK
P
S
P S
Five different Average Noise Level (ANL) conditions; last one showing “No Signal”
Summary of Contents for ACUSON Freestyle
Page 4: ...iv ACUSON Freestyle User Manual ...
Page 14: ...1 2 ACUSON Freestyle User Manual Introduction ...
Page 70: ...3 2 ACUSON Freestyle User Manual Setup ...
Page 78: ...3 10 ACUSON Freestyle User Manual ...
Page 80: ...4 2 ACUSON Freestyle User Manual Patient Studies ...
Page 92: ...5 2 ACUSON Freestyle User Manual Scanning ...
Page 110: ...6 2 ACUSON Freestyle User Manual Wireless Probes ...
Page 122: ...7 2 ACUSON Freestyle User Manual Batteries ...
Page 128: ...8 2 ACUSON Freestyle User Manual Connectivity ...