ACUSON Cypress™ Operator’s Manual
9-19
Transferring Studies
9.
Go to the
Shutdown
tab to view the progress of the transfer or press the
2D
key to exit the menu.
NOTE You can operate the Cypress system while the transfer is in progress
with the exception of the following restrictions:
— You cannot activate a study that has been designated for transfer.
— An active study automatically deactivates when designated for
transfer.
NOTE Studies stored in the
DICOM
format cannot be copied back to the
Cypress system for viewing.
NOTE Segments of the same study can be sent to different destinations in
different formats. For example, an entire study can be sent to the
Cypress internal
MO
disk drive in the Cypress format, selected
(flagged) loops can be sent to a network server, and the report and
selected (flagged) loops can be sent to a computer.
TABLE 43.
Send Screen DICOM Compression Fields
Field
Description
DICOM
Compression
Sets the
DICOM
compression settings. The selected com-
pression is applied to studies saved in
DICOM
format that
are saved to the
MO
disk or saved over a network to a
DICOM
server (if the Network option is installed).
None
: Use when a server or viewer does not support
JPEG
or
RLE
Lossless compression. The largest image
files are produced with this selection.
RLE
Lossless
: Select this format when a lossless data
compression format is desired. This format preserves
the quality of the image. The compression generally
results in a data size reduction of approximately 3:1.
JPEG
Lossy
: A lossy compression format may result in
some degradation of the image, depending on the
compression setting. However,
JPEG
compression
results in significantly smaller file sizes than the
RLE
compression format.
JPEG
Compression
: The
JPEG
Compression scale.
— a
LOWER
setting results in a higher image quality
and larger file sizes.
— a
HIGHER
setting results in lower image quality and