ACUSON Cypress™ Operator’s Manual
5-11
Overview of the Scanning Process
GN #/#/#
Gain
Use the Main knob to adjust the gain. The setting is in
this order:
2D
/
Color
/
Doppler
. For example a setting of:
GN
5/ /13
shows a
2D
gain of 5; no color gain; a Doppler gain
of 13.
For
2D
Gain the range is: 1 through 28.
For Color Gain the range is: 1 through 30.
For Doppler Gain the range is: 1 through 42.
##fps
Frames Per Second
The speed at which the frames are
displayed. Use the
Angle
option on the
2D
image
TOOLS
menu to adjust the rate in 2D mode. Use the
FRAME RATE
Soft Window knob to adjust the rate in color mode.
##.#cm
2D Depth
(2D, Color, and M-Mode only) Use the
DEPTH
Soft Window knob to adjust the depth.
The maximum scanning depth of the system is 26.9 cm.
HPF ###Hz
High Pass Filter
(Doppler Mode only) Use the
FILTER
Soft
Window knob to adjust the filter.
The range is from 200 Hz through 1400 Hz for
CW
and
from 25 Hz through 1400 Hz for
PW
.
F#
Filter
(Color mode only) indicates the velocity filter applied to
the Color Doppler.
The range is F1 through F6.
Harmonic
Harmonic
If your transducer has harmonic mode, the word
“Harmonic” will appear. Use the
FREQUENCY
Soft Window
knob to turn harmonic on or off.
The mode is
ON
when the word
Harmonic
is displayed.
Freq#
Frequency
The frequency of some transducers can be
adjusted. If your transducer can be adjusted the word “Fre-
quency” will be displayed. Use the
FREQUENCY
Soft Window
knob to adjust the frequency.
Invert
Invert
Displayed when the spectral doppler flow display has
been inverted. The spectrum is inverted using the corre-
sponding Soft Window knob.
TABLE 14.
Explanation of the System Data Fields
Setting
Function/Description