ACUSON Cypress™ Operator’s Manual
7-11
Using the Vascular Calculations Package
Deleting a General Measurement
To delete a general measurement, follow these steps:
1.
Display the frame containing the measurement graphic.
2.
Press
CALC
to display the Selection Menu.
3.
Use the Trackball to highlight the measurement you want to delete.
4.
Press the <
BACKSPACE
>> key.
Deleting a Specific Measurement
To delete a specific measurement, follow these steps:
1.
Display the frame containing the measurement graphic.
2.
Press
CALC
to display the Selection Menu.
3.
Use the Trackball to select the name of the Report Menu containing the spe-
cific measurement you want to delete and press <
ENTER
>.
NOTE Specific measurements cannot be deleted from the Selection Menu.
4.
To delete a single measurement, use the Trackball to highlight the measure-
ment you want to delete and press <
BACKSPACE
>. To delete all of the mea-
surements, click [
CLEAR
].
NOTE If the deleted measurement is part of a formula, all the formula results
that use that measurement are also deleted.
Making Corrections to the Trace and Area Measurements
No matter which method you choose to use to perform the measurements, cor-
recting the
Trace
or
Area
outline is the same. Press the <
BACKSPACE
> key to
erase a small portion of the outline. Press and hold the <
BACKSPACE
> key to
continuously erase the path of the cursor until the <
BACKSPACE
> key is
released.