![Siemens ACUSON Cypress Operator'S Manual Download Page 190](http://html.mh-extra.com/html/siemens/acuson-cypress/acuson-cypress_operators-manual_381579190.webp)
Stress Echo
8-2
ACUSON Cypress™ Operator’s Manual
Connecting and Using
ECG
Use
ECG
to display electrocardiogram waveforms. Siemens provides an
ECG
adapter that connects into the back of the Cypress system.
WARNING The Cypress system is not intended for use as an
ECG
monitor or
for diagnostic
ECG
. Signals displayed are for reference or
triggering only.
WARNING Do not use
ECG
patient cables with detachable lead wires that
have exposed male pins. Electrocution could result if these pins
are plugged into
AC
power.
To connect the Cypress system to a stress system, follow these steps:
1.
Assemble the supplied
ECG
-to-
BNC
connector, cable, and
BNC
-to-Stress
System plug as shown in Figure 36, “ECG-to-BNC Connector and BNC-to-
Stress System Plugs,” on page 8-4 and Figure 37, “Stress Echo ECG Cable,”
on page 8-5.
2.
Insert the
ECG
-to-
BNC
connector into the port on the back of the Cypress
system.
3.
Insert the
BNC
-to-Stress System plug into your stress system.
4.
Attach the electrodes and
ECG
leads to the patient.
ECG CABLE
CONNECTIONS
SUPPLIED BY
SIEMENS
(SEE Figure 36 on
CONNECTS
TO YOUR
STRESS
SYSTEM
– o