112
Operating Instructions Combics Indicators
Data Interfaces
Error Messages
Errors are divided into the following:
– Dynamic errors are displayed for the duration of the error with an error code (e.g.
inf 01
).
– Temporary errors are displayed for 2 seconds (e.g.
inf 07
)
– Fatal errors are displayed continuously (e.g.
err101
, a reset is required to clear them).
Display
Cause
Solution
N
o display
No power present
Check power supply
segments
-------
Key has no function in this status
Flashing
k
Battery defective or time changed
Set time
H
Weighing range exceeded
Unload the scale
l
or
Err 54
Weighing pan is not in place
Position the weighing pan
Err 101 - 104
Key is stuck
Release key or key pressed when switching on the device
Contact your local Sartorius Service Center
Err 320
Operating program memory faulty
Contact your local Sartorius Service Center
Err 335
Verifi
ed weighing platform not compatible with the
Connect a compatible weighing platform
connected terminal
Err 340
New EEPROM loaded (Service)
Turn the scale off and then on again. If the error code Err340 is
still displayed, please contact your local Sartorius Service Center
Err 341
RAM has lost data; battery is dead
Leave the scale connected to power for at least 10 hrs.
Err 343
Loss of data in the memory area for transaction numbers
Contact your local Sartorius Service Center
External alibi memory
Inf 01
Data output not compatible with output format
Set output format correctly
inf 02
Adjustment condition was not met
Do not carry out adjustment until after 0 display
e.g. not tared or weighing pan loaded
Unload scale, tare using the
)
key
inf 03
Adjustment could not be completed within a certain time
Allow to warm up again and repeat the adjustment process
Inf 06
Built-in calibration weight defective
Contact your local Sartorius Service Center
inf 07
Function not allowed in scales
verifi
ed
for use in legal metrology
Contact your local Sartorius Service Center
inf 08
The load on the scale is too heavy to zero the readout
Check whether “Tare/zero at power on” (1.12) is set
Inf 09
Taring is not possible when the scale gross weight is < zero
Zero the scale
inf 10
Tare key is blocked when there is data in the tare memory
The data stored for the application program must be deleted
before taring (Combics 2 only).
inf 18
Preload is too light
Inf 19
Preload is too heavy
inf 29
Minimum load not reached
Reduce min. load
(under Application, menu item 3.6)
Inf 30
BPI ID (BPI byte) not deleted in current weighing platform
Reset weighing parameters to factory settings for current
(COM1 is
fi
xed on XBPI data communication)
weighing
platform
Inf 31
Interface handshake interrupted (XOFF, CTS)
Send XON, unblock CTS
inf 71
Cannot store the current weight value
None
(e.g. control limits too low or too high)
inf 72
Cannot store the current weight value
None
(e.g. transaction counter maximum reached)
inf 73
Data cannot be written or read
Contact your local Sartorius Service Center
inf 74
Function is blocked (e.g. menu is locked, device is already
None
confi
gured to another interface)
NO WP
No weighing platform connected
Connect a weighing platform
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...