83
–
Value saved automatically
when the weighing platform is stable and the
defi
ned minimum load is exceeded.
If the
defi
ned minimum load is not exceeded, you can save the item
manually
by pressing the
O
key. Regardless of these settings, the current value cannot
be saved automatically unless the platform is unloaded before the next sample is
placed on it. The weighing platform is considered to be unloaded when the load
is less than 50% of the minimum load.
The number of items added to memory is shown in the number display.
Press the
c
key to clear the totalizing memory. A printout is automatically
generated.
With two weighing platforms connected, you can add values from both platforms
to the totalizing memory. The displayed result is accurately calculated in the active
weight unit.
Example: When you add 1.243 g (determined on a weighing platform with three
decimal places) to 1.4 g (determined on a platform with 1 decimal place), the display
shows 2.643 kg.
Preparation
t
Open the
APPLIC. / APPLIC.3 / TOTALIZ
menu.
Available Parameter Settings
* = Factory setting
MIN.INIT
Minimum load for initialization
3.6
1 digit
1 digit
3.6.1*
2 digit
2 digits
3.6.2
5 digit
5 digits
3.6.3
10 digT.
10 digits
3.6.4
20 digT.
20 digits
3.6.5
50 digT.
50 digits
3.6.6
100 diG.
100 digits
3.6.7
200 diG.
200 digits
3.6.8
500 diG.
500 digits
3.6.9
1000 dI.
1000 digits
3.6.10
AUTO.SAV
Autosave
3.16.
OFF
Off
3.16.1*
ON
On
3.16.2
PRT.SAV.
Individual/Component printout when saved
3.17.
OFF
Automatic printing off
3.17.1
ON
Print the entire standard print
confi
guration
every time with the
O
key is pressed
3.17.2*
VAL.FROM
Source of data for autosave
3.22.
APPLIC.1
Application 1
3.22.1*
APPLIC.2
Application 2
3.22.2
SAV.VAL.
Save value
3.23.
NET
Net
3.23.1*
CALCUL.
Calculation
3.23.2
NET+CAL
Net and Calculated
3.23.3
Printout
You can
confi
gure whether a printout is generated automatically when a weight
value is stored in the totalizing memory or manually by pressing the
p
key.
Setting:
Applic./Applic.3/TOTALIZ
menu item 3.17.
-
You can print manually by pressing the
p
key (single printout):
3.17.1
-
Component log (single printout of an item):
3.17. 2
The total data record is printed when you clear the totalizing memory (by pressing
the
c
key).
Operation
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...