62
Operating Instructions Combics Indicators
Minimum Load for Initialization
You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform is
too light, the following will occur:
– Error code
Inf 29
appears,
– The weighing platform is not initialized,
– The preset reference sample quantity is saved.
Setting:
Applic./Applic.1/ANIM.WG/Min.init
menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The “digits” here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000
digits are required, you must place at least 1000 g (= 1000 intervals =1000 digits)
on the weighing platform for initialization.
Starting the Measurements
The averaging routine does not begin until the fl uctuation in weight value remains
below a
defi
ned threshold over three consecutive measurements. The tolerance limit
is
defi
ned as a percentage of the animal or object weight (for example, 0.1%, 0.2%,
..., 50%, 100%),
confi
gured in Setup under:
Activity
menu item 3.19.
If the “Averaging” parameter is set to 2%, for example, and the animal or object
weighs 10 kg, measurement does not begin until the fl uctuation in weight value
remains below 200 g during three consecutive measurements.
Display
A calculated average value is shown continuously on the main display. The
G
symbol
indicates the calculated value.
You can toggle between the results display and the current scale display by pressing
the
w
key.
Setting:
Applic./Applic.1/ANIM.WG/DIS.UNLD
menu item 3.21.
You can select “Display is static until unload threshold reached” to have the display
switch automatically to the weight readout when you unload the weighing platform
(i.e., when the load is less than half the minimum load). The result of the most recent
averaging operation is not saved.
If you select “Display is static until the
c
key is pressed,” the calculated average
remains displayed even after the weighing platform is unloaded, until you press the
c
key to begin a new measurement.
Example
: The weight of one mouse should be measured.
Confi
guration: The “Animal Weighing” application is selected, and printout has been
set up (see
“Confi
guration”).
Operation
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...