Operating Instructions Combics Indicators
27
Adjustment without Weights
In the Service menu, adjustment without weights can be carried out by entering the
characteristic data of the load cells.
3
Adjustment without weights may not be carried out on weighing equipment used
in legal metrology.
Setup Information
– Adjustment without weights is only possible when the menu access switch is
open in the Service menu.
– When the Service mode is activated, you can enter the necessary parameters for
adjustment without weights in the
Setup
menu under
WP-1
for the
fi
rst
weighing platform and
COm1 / WP-2
,
UNICOM / WP-2
or
COm-WP
for the second
weighing platform. The settings are made in the corresponding Setup menu
under menu item 1.19.
– The “Nominal load” parameter must be entered in the kg unit.
– The “Resolution” parameter must be entered in the kg unit and must correspond
to the scale interval d entered for the ADC
confi
guration.
– The “Sensitivity” parameter is entered in mV/V ( see the data sheet for the
value).
h
The data entered are saved by selecting menu item 1.19.7. After saving, the data will
no longer be able to be read.
Procedure
t
Open menu access switch.
t
Activate the Service mode.
t
Select the weighing platform.
t
Enter the nominal load of the load cell(s) in kg under menu item 1.19.1. If the
weighing platform has multiple load cells, the nominal load must be multiplied
accordingly (e. g. 4 load cells, each of which has a capacity of 50 kg, will
produce a nominal load of 200 kg).
t
Enter the resolution in kg under menu item 1.19.2. The value must correspond
to the scale interval d entered under menu item 11.4.1.
t
Enter the sensitivity of the load cells in mV/V under menu item 1.19.3.
For weighing platforms with multiple load cells: Enter the individual values of
the load cells in 1.19.3 to 1.19.6 or enter the average of all load cells in 1.19.3.
t
Save the values for adjustment without weighing under menu item 1.19.7.
t
Close the menu access switch.
Menu structure for adjustment without weights
ADJ.W/O.W
Adjustment without weights (entering the characteristic data
of the load cell(s))
1.19.
NOM.LOAD Nominal load
1.19.1
RESOLUT Resolution
1.19.2
SENSIT.1
Sensitivity in mV/V for cell 1 (or average value for all load cells)
1.19.3
SENSIT.2
Sensitivity in mV/V for cell 2
1.19.4
SENSIT.3
Sensitivity in mV/V for cell 3
1.19.5
SENSIT.4
Sensitivity in mV/V for cell 4
1.19.6
SAVE
Save values for 1. 19
1.19.7
Confi guring Weighing Platforms
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...