52
Operating Instructions Combics Indicators
Entering the reference piece weight
The reference piece weight (i.e., the weight of one piece) can be entered using the
keypad and save with the
O
key.
Reading the reference piece weight
The reference piece weight can be read using a barcode scanner.
h
The entered value remains active until deleted by pressing the
c
key or until
overwritten by a new value. It remains saved after the scale is switched off.
Preparation
t
Open the
APPLIC. / APPLIC.1 / COUNT.
menu.
Available Parameter Settings
* = Factory setting
MIN.INIT
Minimum load for initialization
3.6
1 digit
1 digit*
3.6.1
2 digit
2 digits
3.6.2
5 digit
5 digits
3.6.3
10 digT.
10 digits
3.6.4
20 digT.
20 digits
3.6.5
50 digT.
50 digits
3.6.6
100 diG.
100 digits
3.6.7
200 diG.
200 digits
3.6.8
500 diG.
500 digits
3.6.9
1000 dI.
1000 digits
3.6.10
RESOLUT
Resolution for calculation of reference value
3.9.
DISP.ACC
Display accuracy*
3.9.1
10 FOLD
Display ac 1 decimal place
3.9.2
100FOLD
Display ac 2 decimal places
3.9.3
SAVE WT.
Parameter for saving weight values
3.11
STABIL.
With stability*
3.11.1
ACC.STAB
With increased stability
3.11.2
REF.UPDT
Reference sample updating
.
3.12.
OFF
Off
3.12.1
automat
Automatic*
3.12.3
REF.WP
Reference weighing instrument
3.13.
NO WP
No weighing platform selected
3.13.1*
wp 1
Weighing platform WP1
3.13. 2
wp 2
Weighing platform WP2
3.13. 3
)
�
t
To save the setting, press the
)
key.
( (
...
t
To exit setup: Press the
(
key several times.
Minimum Load for Initialization
You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
– Error code
Inf 29
appears,
– The weighing platform is not initialized,
– The preset reference sample quantity is saved.
Setting:
Applic./Applic.1/Count./Min.init
menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The “digits” here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000 digits
are required, you must place at least 1000 g (= 1000 intervals = 1000 digits) on the
weighing platform for initialization.
Operation
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...