Operating Instructions Combics Indicators
45
Features
Which of the following features are available depends on the connected weighing
platform. These features are
confi
gured in the
Setup
menu:
– External calibration/adjustment blocked in
verifi
ed weighing instruments
– External calibration/adjustment with the default weight value or
user-defi
ned
weight (not available on
verifi
ed instruments):
setup / wp-1/ param.1 /CAL./ADJ.
menu “Calibration, Adjustment."
– Specify the weight for external calibration/adjustment:
setup / wp-1
/ param.1 / man.ext.w
menu “Entering Calibration Weight"
– Internal adjustment for IS weighing platforms
(confi
gure under:
COM1
or
unicom / WP2
), Combics 2
– Block the
J
key to prevent use of the two functions described above:
setup / wp-1
“Calibration, Adjustment."
– Calibrate
fi
rst; then adjust automatically or manually (not for
verifi
ed weighing
instruments):
setup / wp-1/ param.1 / Cal.routine
"Calibration/Adjustment
sequence"
– Flashing
n
symbol as adjustment prompt. If more than one weighing
platform is connected, the platform number is also displayed:
setup / wp-1
menu Menu item 1.15. “Adjustment prompt"
– Block external calibration/adjustment:
setup / wp-1/ param.1 / EXT.Cal.
menu
“External calibration."
Example:
External Calibration and Manual Adjustment with Default Weights
(weighing
parameters: factory settings)
Operation
(
1.) Zero the scale.
J
2.) Start calibration
(e.g., when adjustment
prompt fl ashes WP).
cal.Ext.
is displayed for
two seconds.
You are prompted to place
the required weight on the
platform (e.g., 10,000 g).
3.) Position the calibration/
adjustment weight on the
weighing platform.
The difference between the
measured value and the
true weight of the sample
will be displayed with plus/
minus signs.
Ext. calibration
Printout is generated,
Nom. + 10000 g
if the process was cancelled
Diff. + 1 g
with
(
J
4.) Activate calibration/
adjustment (press the
(
key to cancel).
The adjustment weight is
displayed once adjustment
is
fi
nished.
--------------------
A GMP-compliant printout
24.02.2010 10:15
is generated.
Type CAW2S
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-02
--------------------
Ext. calibration
Nom. + 10000 g
Diff. + 1 g
Ext. adjustment
Diff. + 0 g
--------------------
24.02.2010 10:15
Name:
--------------------
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...