100
Operating Instructions Combics Indicators
Data Interfaces
Analog UniCOM Interface (optional)
Level:
4 to 20 mA, 0 to 20 mA, 0 to 24 mA, 0 to 10V
Power supply:
Internal
Factory setting:
4 to 20 mA
Connection:
CAWxP, CAISL2 devices (IP44 protection):
25-pin D-Sub socket
CAWxS, CAH1 devices (IP69K protection):
Connection via screw terminals in the housing,
cable routed into the housing via a cable gland.
1)
Optional UniCOM universal data interface
2)
XBPI operating mode: 9600 baud, 8 data bits, parity: odd, 1 stop bit
3)
Only with the standard COM1 interface
4)
Network address is valid only in XBPI-RS485 operating mode
Connection Options
The following
printers
can be connected to
COM1 and UniCOM
– YDP20
(user-defi
ned interface parameters)
– YDP14IS (strip or label printer)
– YDP04IS (strip or label printer)
– Universal printer
(user-defi
ned transmission parameters)
The following devices can also be connected to
COM1
:
– Foot / Hand switch
– PC (RS-232 interface)
– 2nd weighing platform (Combics 2 only, RS-232 interface)
– External checkweighing display (stop light) via a digital I/O (Sartorius standard)
The following devices can also be connected to the
optional
UniCOM
:
– PC (RS-232 interface)
– 2nd weighing platform (Combics 2 only, in RS-232 or RS-485 mode)
– 2nd printer (external power source required)
– Remote display
– Current interface (4-20 mA, 0-20 mA, 0-24 mA, 0-10 V)
–
Profi
bus
– Ethernet
– Galvanically isolated digital I/0
3
You may need to use an external power supply to operate peripheral devices.
Connecting a
second weighing platform
:
A second weighing platform can be connected to the Combics 2 at COM1, UniCOM
or COM-WP.
COM1
is operated in the RS-232 mode. A second weighing platform can use the
following operating modes:
– SBI standard
– SBI
verifi
able
– XBPI-232 (factory setting)
– ADC-232
UniCOM
can be operated in either RS-232 or RS-485 mode. A second weighing
platform can use the following operating modes:
– SBI (RS-232 mode)
– XBPI-232 (RS-232 mode)
– ADC-232 (RS-232 mode)
– IS-485 (RS-485 mode, XBPI mode, factory setting)
– ADC-485 (RS-485 mode)
Summary of Contents for Combics CAH1
Page 126: ...126 Operating Instructions Combics Indicators Declarations of Conformity...
Page 127: ...Operating Instructions Combics Indicators 127 Declarations of Conformity...
Page 128: ...128 Operating Instructions Combics Indicators EC Type approval certificate...
Page 129: ...Operating Instructions Combics Indicators 129 certificate...
Page 130: ...130 Operating Instructions Combics Indicators Test Certificate...
Page 131: ...Operating Instructions Combics Indicators 131 Test Certificate...
Page 132: ...132 Operating Instructions Combics Indicators Test Certificate...
Page 133: ...Operating Instructions Combics Indicators 133 Plates and Markings...
Page 134: ...134 Operating Instructions Combics Indicators Plates and Markings...
Page 135: ...Operating Instructions Combics Indicators 135 Plates and Markings...
Page 136: ...136 Operating Instructions Combics Indicators Plates and Markings...
Page 137: ...Operating Instructions Combics Indicators 137 Plates and Markings...
Page 138: ...138 Operating Instructions Combics Indicators Plates and Markings...
Page 139: ...Operating Instructions Combics Indicators 139 Certificate...
Page 166: ...166 Operating Instructions Combics Indicators...
Page 167: ...Operating Instructions Combics Indicators 167...