4-27:
Fig.4-27 Magnitude-frequency characteristics diagram of 2.3GHz-2.4GHz band-pass filter
10 Coverage map (option)
The interference map option can be used for RSSI testing and adjacent channel power ratio testing, and the test
results can be marked on the map in real time based on time or distance. The test results marked on the map can
be saved to the instrument for later recall.
1) RSSI measurement
Main operation steps of RSSI measurement:
a) Press [Measure]→[Coverage Map]→[Coverage Map Off On], in the default, the measurement mode is RSSI
measurement after the interference map switch is turned on.
b) Press [Freq]→[Center Freq], set the measured signal frequency.
c) Press [Measure]→[Coverage Map]→[Measurement Setup]→[RSSI], set the color of the different signals
received on the map.
d) Press [Measure]→[Coverage Map]→[Distance Time]→[Repeat Type Time Dist], set the way to map the
information according to the time interval or how to move the distance.