Fig. 9-1 Schematic Diagram of PScan Mode
2
FScan
Main operation steps of FScan:
1) Press [Measure]→ [FScan] to enable the frequency sweeping mode.
2) Press [Freq]→ [Start Freq] and set the starting frequency of sweeping.
3) Press [Freq]→ [Step Freq] and set the step frequency of sweeping.
4) Press [Freq]→ [Points] and set the number of sweeping points.
5) Press [BW]→ [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz,
50kHz, 120kHz or 150kHz.
6) Press [Swp/Ant]→ [Recall Antenna], and select the antenna factor file.
7) Press [Swp/Ant]→ [Dwell Time +∞ Man], and set the dwell time.
8) Press [Swp/Ant]→ [Dwell Time Off On] to enable or disable the dwell time.
9) Press [Marker]→ [Marker Off On] to enable or disable the marker.